Can venous saturations from the central venous line and the venous side of the heart–lung machine be interchangeable with mixed venous saturation from the pulmonary artery in children undergoing open-heart surgery?


1 Anesthesia Department, College of Medicine, Assiut University, Assiut, Egypt
2 Cardiothoracic Surgery Department, College of Medicine, Assiut University, Assiut, Egypt

Date of Submission29-Mar-2017
Date of Acceptance09-Jan-2018
Date of Web Publication9-Feb-2018

Correspondence Address:
Dr. Sayed K Abd-Elshafy
Department of anesthesia, Assiut University Hospital, Assiut
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ejca.ejca_8_17

Rights and Permissions
  Abstract 

Background Mixed venous oxygen saturation has been advocated as an indirect index of tissue oxygenation. This study evaluated whether venous oxygen saturation from the central venous line (CVL) or the venous side of the cardiopulmonary bypass (CPB) was interchangeable with mixed venous oxygen saturation from the pulmonary artery in children undergoing cardiac surgery for correction of congenital heart defects.
Patients and methods Forty children ranging in age from 1 to 15 years undergoing correction for congenital heart procedures with CPB were included in this study. Simultaneous samples were taken from the CVL, the pulmonary artery, and from the venous side of the CPB after 10 min on full CPB and at the end of surgery before weaning from CPB. Bland and Altman’s analysis was carried out to study the agreement between different venous oxygen saturation.
Results Insignificant correlations were observed between venous oxygen saturation from the pulmonary artery, the CVL, and the venous site of the CPB. Wide limits of agreements were observed between the venous oxygen saturation in the pulmonary artery with that in the CVL (14.21–−15.32), and also with that in the venous side of the CPB (34.34–−33.18). A wide limit of agreement was observed between venous oxygen saturation in the venous side of the CPB with that in the CVL (28.24–−31.67).
Conclusion Venous saturations from the CVL, and from venous side of CPB are not interchangeable with mixed venous saturation from the pulmonary artery in children undergoing open-heart surgery.

Keywords: children, congenital, defects, heart, oxygen, venous


How to cite this article:
Ali MS, Abd-Elshafy SK, Abd Allah EM, Ghoneim AF. Can venous saturations from the central venous line and the venous side of the heart–lung machine be interchangeable with mixed venous saturation from the pulmonary artery in children undergoing open-heart surgery?. Egypt J Cardiothorac Anesth 2017;11:48-52

How to cite this URL:
Ali MS, Abd-Elshafy SK, Abd Allah EM, Ghoneim AF. Can venous saturations from the central venous line and the venous side of the heart–lung machine be interchangeable with mixed venous saturation from the pulmonary artery in children undergoing open-heart surgery?. Egypt J Cardiothorac Anesth [serial online] 2017 [cited 2018 Apr 20];11:48-52. Available from: http://www.ejca.eg.net/text.asp?2017/11/3/48/225084


  Introduction Top


Mixed venous oxygen saturation (SvO2) refers to the hemoglobin oxygen saturation of blood drawn from the proximal pulmonary artery and reflects the oxygen balance of the entire body [1],[2],[3]. SvO2 was proposed as an indicator of the balance between systemic oxygen delivery and oxygen consumption and thus can be used to assess the adequacy of tissue oxygenation [1],[2],[4]. Mixed venous oxygen saturation monitoring requires the introduction of a pulmonary artery catheter which carried significant complications [5],[6]. Venous oxygen saturation from the internal jugular vein (ScvO2) refers to a hemoglobin oxygen saturation of blood from the superior vena cava [2],[3],[4]. This measurement reflects the venous oxygen saturation of blood from the brain and the upper body but neglects venous blood from the lower body [2],[7],[8],[9],[10]. Monitoring of the ScvO2 was suggested as safer and more cost-effective than the SvO2 [7],[8],[9],[10]. Also, insertion of the central venous line (CVL) is used routinely in adults and children undergoing open-heart surgery with cardiopulmonary bypass (CPB). However, there has been a major debate in the recent literature on the use of ScvO2 as an alternative to SvO2 in septic patients or during different surgical procedures [1],[2],[3],[4]. During cardiac surgery, venous oxygen saturation from the venous side of the heart–lung machine has been used to express mixed venous oxygen saturation [11],[12],[13]. There is little appeared in the literature on the relationship between mixed venous oxygen saturation from the pulmonary artery, and venous oxygen saturation from the CVL and that from the venous side of the heart–lung machine in children undergoing open-heart surgery for correction of congenital heart defects. This study aimed to examine the relationship between venous oxygen saturation from the CVL and the venous side of the heart–lung machine and also their interchangeability with mixed venous oxygen saturation from the pulmonary artery in children undergoing open-heart surgery.


  Patients and methods Top


The study is a prospective one and was approved by the Local Ethics Committee of the Faculty of Medicine, Assiut University and was registered with ClinicalTrials.gov (Clinicaltrial.gov identifier: NCT02498626). A written parents’ informed consent was obtained. Forty children ranging in the age from 1 to 15 years and scheduled for surgical closure of the different congenital heart defects with CPB in the period from July 2009 until January 2016 were studied. Patients with hepatic or renal impairment, and emergency surgery or reoperation were excluded. Preoperative preparation, induction, and maintenance of anesthesia, patient monitoring, and a CPB technique have been described in a previous paper [11]. A radial artery catheter and central venous catheter were inserted after induction of anesthesia. Two-lumen indwelling catheters were used for central venous catheterization (double-lumen 5 Fr ARROW Blue Catheters, 8 or 13 cm, Blue FlexTip; B. Braun Melsungen AG, Melsungen, Germany). Chest radiography was performed to confirm the position of the internal jugular catheter in the superior vena cava in all cases.

Study measurements

Samples from three sites for venous oxygen saturation (pulmonary artery by the surgeon through direct access by a very small needle, internal jugular vein, and from the venous blood of the CPB) were obtained 10 min after full-flow CPB and another sample was obtained before weaning from CPB. Samples were analyzed using a blood gas analyzer (EDAN i15-FCC: SMQi15EDAN-China).

Statistical analysis

Sample size calculation was based on a power of 80% to detect differences between pairs of measurements and an α error of 5%. This yielded a sample of 72 pairs of measurements in 36 patients, and 80 pairs of measurements were included from 40 patients to compensate for any dropouts during the study. All results were analyzed using SPSS version 20 for Windows (SPSS Inc., Chicago, Illinois, USA). Correlations and Bland and Altman analysis with limits of agreement 14 were carried out to study the relationship between venous saturations from the three sites. Mean difference (bias), a SD of this difference (precision), confidence interval, and limits of agreement between venous saturation measurements were reported.


  Results Top


Forty children who underwent correction of different congenital heart diseases were studied. Patient characteristics are presented in [Table 1]. Various congenital heart defects are shown in [Table 2]. No significant correlations were observed between venous oxygen saturation in the pulmonary artery, in the CVL, and in the venous site of the heart–lung machine ([Table 3]). A wide limit of agreements occured between venous oxygen saturation in the pulmonary artery with that in the CVL (14.21–−15.32), and also with that in the venous side of the heart–lung machine (34.34–−33.18) ([Table 3] and [Figure 1],[Figure 2],[Figure 3]). Also, a wide limit of agreement was observed after analyzing data from patients with noncyanotic lesions (23 patients) and those with cyanotic lesions (17 patients) ([Table 4] and [Table 5]).
Table 1 Patient characteristics

Click here to view
Table 2 Type of congenital heart lesions

Click here to view
Table 3 Correlations and limits of agreement between venous saturations of the central venous line, pulmonary artery and venous blood of cardiopulmonary bypass (sample size=80)

Click here to view
Table 4 Correlations and limits of agreement between venous saturations of the central venous line, pulmonary artery and venous blood of cardiopulmonary bypass for the patients with noncyanotic lesions (sample size=46)

Click here to view
Table 5 Correlations and limits of agreement between venous saturations of the central venous line, pulmonary artery and venous blood of cardiopulmonary bypass for the patients with cyanotic lesions (sample size=34)

Click here to view
Figure 1 Bland and Altman plot of venous oxygen saturations from the venous side of the heart–lung machine and the central venous line.

Click here to view
Figure 2 Bland and Altman plot of venous oxygen saturations from pulmonary artery and the central venous line.

Click here to view
Figure 3 Bland and Altman plot of venous oxygen saturations from pulmonary artery and the venous side of the heart–lung machine.

Click here to view



  Discussion Top


In this study, venous oxygen saturation from the internal jugular vein (ScvO2) or from that of the venous blood of the CPB is not in agreement with mixed venous oxygen saturation from the pulmonary artery (SvO2). The lack of this agreement was also found on analyzing of the results for cyanotic and noncyanotic lesions. Several studies have used the central venous saturation as an alternative to mixed venous saturation from the pulmonary artery in patients with different medical conditions, and the results are conflicting [3],[14],[15],[16],[17],[18],[19],[20],[21]. Vapula et al. [3] found broad limits of agreement (−8.1–16.5%) between SvO2 and ScvO2 in septic patients. They concluded that a mixed venous oxygen saturation caught not be estimated by ScvO2. Kopterides et al. [15] and Ho et al. [16] reported similar findings. They found that the limits of agreement were −20.2–3.3% and −5.0–18.8; in both studies respectively. The same results were found during and after cardiac surgery in adult patients, where the mean bias between SvO2 and ScvO2 was 4.4%, and the limits of agreement were −13.6–22.3% [17]. The previous reports are in agreement with our study, where limits of agreement were −15.3–14.2%). Such differences and the lack of agreement between SvO2 and ScvO2 could be explained by the blood coming from the pulmonary artery reflects venous saturation throughout the entire body including that from the superior vena cava [2], whereas ScvO2 mainly reflects blood drained from the brain and upper parts of the body [7],[8],[9],[10]. In contrast to our findings, other studies reported that ScvO2 measurements were closely related to SvO2 [18],[19]. Interestingly, in this study, venous oxygen saturation from the venous side of the CPB was not related to SvO2 from the pulmonary artery, as it had limits of agreement ranging from −33.18 to 34.34%. In one animal study, oxygen saturation from the venous side of the CPB did not reflect regional venous oxygen saturation from the portal vein or that from the sagittal sinus. Ten SvO2 did not indicate brain oxygen saturation. Brain oxygen saturation is monitored by jugular bulb oximetry [12] or near-infrared spectroscopy [22] in patients undergoing cardiac surgery with CBP. It is clear that regional oxygen saturation cannot be reflected by monitoring venous oxygen saturation from the heart–lung machine or that from the pulmonary artery [2],[12],[22],[23]. This can be explained in part by the poor agreement between venous oxygen saturation from the internal jugular vein and the venous side of the heart–lung machine with mixed venous oxygen saturation from the pulmonary artery in our study. Moreover, Kakuta et al. [24] found parallel changes between ScvO2 from the internal jugular vein with regional brain oxygen saturation measured by near-infrared spectroscopy (rSO2) despite the marked differences in ScvO2 in patients undergoing a bidirectional Glenn shunt. However, the authors did not use Bland and Ataman [25] analysis to test the agreement between the two measurement tools.

The fact that this study was carried out only during the CPB period and that only intermittent samples were obtained can be considered limitations of this study.


  Conclusion Top


It appears that venous saturations from the CVL and from the venous side of the heart–lung machine cannot substitute for or be interchange with mixed venous saturation from the pulmonary artery in children undergoing open-heart surgery.

Acknowledgements

Equipment used in this study was provided by Assiut University Hospitals, Assiut, Egypt.

Study design performed by M.S., S.K., E.A. and A.G. Patients recruited by E.A. and A.G. Data collected by M.S., S.K., and E.A. Data analyzed by M.S., S.K., and A.G. The first draft was written by M.S., S.K., and E.A. Revising the paper was done by S. K. and A.G.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Bloos F, Reinhart K. Venous oximetry. Intensive Care Med 2005; 31:911–913.  Back to cited text no. 1
    
2.
Van Beest P, Wietasch G, Scheeren T, Spronk P, Kuiper M. Clinical review: use of venous oxygen saturation as a goal − a yet unfinished puzzle. Crit Care 2011; 15:232.  Back to cited text no. 2
    
3.
Varpula M, Karlsson S, Ruokonen E, Pettilä V. Mixed venous oxygen saturation cannot be estimated by central venous oxygen saturation in septic shock. Intensive Care Med 2006; 32:1336–1343.  Back to cited text no. 3
    
4.
Marx G, Reinhart K. Venous oximetry. Curr Opin Crit Care 2006; 12:263–268.  Back to cited text no. 4
    
5.
Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D et al. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomized controlled trial. Lancet 2005; 366:472–477.  Back to cited text no. 5
    
6.
Rajaram SS, Desai NK, Kalra A, Gajera M, Cavanaugh SK, Brampton W et al. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst Rev 2013; 2:CD003408.  Back to cited text no. 6
    
7.
Bracht H, Hänggi M, Jeker B, Wegmüller N, Porta F, Tüller D et al. Incidence of low central venous oxygen saturation during unplanned admissions in a multidisciplinary intensive care unit: an observational study. Crit Care 2007; 11:R2.  Back to cited text no. 7
    
8.
Rivers EP, Ander DS, Powell D. Central venous oxygen saturation monitoring in the critically ill patient. Curr Opin Crit Care 2001; 7:204–211.  Back to cited text no. 8
    
9.
Turnaoğlu S, Tuğrul M, Camci E, Cakar N, Akinci O, Ergin P. Clinical applicability of the substitution of mixed venous oxygen saturation with central venous oxygen saturation. J Cardiothorac Vasc Anesth 2001; 15:574–579.  Back to cited text no. 9
    
10.
Ramakrishna M, Hegde D, Kumaraswamy G, Gupta R, Girish T. Correlation of mixed venous and central venous oxygen saturation and its relation to a cardiac index. Indian J Crit Care Med 2006; 10:230–234.  Back to cited text no. 10
  [Full text]  
11.
Yazigi A, El Khoury C, Jebara S, Haddad F, Hayeck G, Sleilaty G. Comparison of central venous to mixed venous oxygen saturation in patients with low cardiac index and filling pressures after coronary artery surgery. J Cardiothorac Vasc Anesth 2008; 22:77–83.  Back to cited text no. 11
    
12.
McDaniel LB, Zwischenberger JB, Vertrees RA, Nutt L, Uchida T, Nguyen T et al. Mixed venous oxygen saturation during cardiopulmonary bypass poorly predicts regional venous saturation. Anesth Analg 1995; 80:466–472.  Back to cited text no. 12
    
13.
Croughwell ND, White WD, Smith LR, Davis RD, Glower DD Jr, Reves JG et al. Jugular bulb saturation and mixed venous saturation during cardiopulmonary bypass. J Card Surg 1995; 10(Suppl):503–508.  Back to cited text no. 13
    
14.
Van Beest PA, van Ingen J, Boerma EC, Holman ND, Groen H, Koopmans M et al. No agreement of mixed venous and central venous saturation in sepsis, independent of sepsis origin. Crit Care 2010; 14:R219.  Back to cited text no. 14
    
15.
Kopterides P, Bonovas S, Mavrou I, Kostadima E, Zakynthinos E, Armaganidis A. Venous oxygen saturation and lactate gradient from superior vena cava to pulmonary artery in patients with septic shock. Shock 2008; 31:561–567.  Back to cited text no. 15
    
16.
Ho KM, Harding R, Chamberlain J, Bulsara M. A comparison of central and mixed venous oxygen saturation in circulatory failure. J Cardiothorac Vasc Anesth 2010; 24:434–439.  Back to cited text no. 16
    
17.
Romagnoli S, Ricci Z, Balsorano P, Tujjar O, Villa G, Bevilacqua S et al. Comparison between mixed and central venous oxygen saturation in patients with severe acute heart failure after cardiac surgery: a prospective observational study. Int J Cardiol 2014; 175:566–567.  Back to cited text no. 17
    
18.
Tahvanainen J, Meretoja O, Nikki P. Can central venous blood replace mixed venous blood samples? Crit Care Med 1982; 10:758–761.  Back to cited text no. 18
    
19.
Ladakis C, Myrianthefs P, Karabinis A, Karatzas G, Dosios T, Fildissis G et al. Central venous and mixed venous oxygen saturation in critically ill patients. Respiration 2001; 68:279–285.  Back to cited text no. 19
    
20.
Dueck MH, Klimek M, Appenrodt S, Weigand C, Boerner U. Trends but not individual values of central venous oxygen saturation agree with mixed venous oxygen saturation during varying hemodynamic conditions. Anesthesiology 2005; 103:249–257.  Back to cited text no. 20
    
21.
Reinhart K, Kuhn HJ, Hartog C, Bredle DL. Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill. Intensive Care Med 2004; 30:1572–1578.  Back to cited text no. 21
    
22.
Dullenkopf A, Baulig W, Weiss M, Schmid ER. Cerebral near-infrared spectroscopy in adult patients after cardiac surgery is not useful for monitoring absolute values but may reflect trends in venous oxygenation under clinical conditions. J Cardiothorac Vasc Anesth 2007; 21:535–539.  Back to cited text no. 22
    
23.
Moreno GE, Pilán ML, Manara C, Magliola R, Vassallo JC, Balestrini M et al. Regional venous oxygen saturation versus mixed venous saturation after pediatric cardiac surgery. Acta Anaesthesiol Scand 2013; 57:373–379.  Back to cited text no. 23
    
24.
Kakuta N, Kawahito S, Mita N, Kambe N, Kasai A, Wakamatsu N et al. Usefulness of central venous oxygen saturation monitoring during bidirectional Glenn shunt. J Med Invest 2013; 60:272–275.  Back to cited text no. 24
    
25.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1:307–310.  Back to cited text no. 25
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Patients and methods
Results
Discussion
Conclusion
References